74
Reference Manual
00809-0100-4811, Rev DA
Section 4: Basic Configuration/Start-Up
January 2015
Basic Configuration/Start-Up
4.5 Volume configuration
4.5.1 Transmitter variables
Select the Volume option to configure the transmitter for volume measurements.
4.5.2 Volume units
Select one of the following units:
Gallons
Liters
Imperial Gallons
Cubic Meters
Barrels
Cubic Yards
Cubic Feet
Cubic Inch
4.5.3 Tank type
Select a standard tank shape, or select the strapping option. Standard shapes are: Vertical
Cylinder, Horizontal Cylinder, Vertical Bullet, Horizontal Bullet or Sphere. (If the PV is Level,
select None for Tank Type).
If your tank does not correspond to any of the above tank shapes, select Strap Table.
4.5.4 Tank dimensions
If a standard tank type was chosen, enter the diameter and height of the tank. See Volume
configuration on page 64 for information on how to specify tank dimensions.
4.5.5 Strapping table
If tank type Strapping Table was chosen, enter how many entries you will use, and the actual
level and volume points. You can enter from 2 to 10 points. The strapping points must be
entered such that the first point corresponds to the lowest level, and the last point corresponds
to the topmost level of the tank.
HART Comm
1, 1, 1, 1
HART Comm
1, 3, 1, 2
HART Comm
1, 4, 3, 1
HART Comm
1, 4, 3, 2-3
HART Comm
1, 4, 3, 4