110
Reference Manual
00809-0100-4811, Rev DA
Section 6: Service and Troubleshooting
January 2015
Service and Troubleshooting
6.3.7 Removing the transmitter head
Note
See the safety information on page 93 before removing the transmitter head.
1. Loosen the nut that connects the
transmitter housing to the process seal.
2. Lift the transmitter head carefully.
3. Make sure the upper surface of the
process seal is clean and the spring-loaded
pin at the center of the process seal is
properly inserted (the pin should move
back when pushed into the hole).
4. Attach the protection plug to the process
seal.
Note
Do not remove the process seal from the
adapter!
Nut
Process seal
Put the protection
plug here
FLANGE VERSION
Nut
Process seal
Adapter
THREADED VERSION
Put the protection
plug here
Komentarze do niniejszej Instrukcji