Astar HT-3300 Dokumentacja Strona 117

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 116
107
Reference Manual
00809-0100-4811, Rev DA
Section 6: Service and Troubleshooting
January 2015
Service and Troubleshooting
6.3.5 Logging measurement data
To start logging, do the following:
1. Select the Monitor icon in the RCT workspace or select the Monitor option from the
View menu.
2. Select the desired variables to be monitored and select the Start Monitor
button.
Saving the log to disk
1. Select the desired variables to be monitored.
2. Select the Log interval button and enter a time interval. For example, type 10 if
you want data to be logged in every tenth second.
3. Select the Counter button and enter the maximum number of files to be stored. The
Counter is used to limit the amount of data stored on the hard disk. Each time the
maximum number of entries in a log file is reached, the current log file is saved and a
new file is created. This procedure continues up to the maximum number of files given
by the counter value. The file size is limited to 60,000 entries which can easily be
handled by spreadsheet programs like MS Excel.
4. Select the desired options for timer, time, and date. By selecting a check-box, the
corresponding time indication is stored for each log entry in the log file.
5. Select the Start disk logging button.
6. Select a destination folder and enter a file name.
Start disk logging
Start monitoring CounterLog interval