48
Reference Manual
00809-0100-4811, Rev DA
Section 3: Installation
January 2015
Installation
Coaxial probe
The coaxial probe can be guided by a tube welded
on the tank bottom. Tubes are customer
supplied. Make sure the probe can move freely in
order to handle thermal expansion.
Rigid twin lead probe
The Rigid Twin Lead probe can be secured to the
tank wall by cutting the center rod and putting a
fixture at the end of the outer rod.
The fixture is customer supplied. Make sure the
probe is only guided and not fastened in the
fixture to be able to move freely for thermal
expansion.
Flexible single lead probe
The probe rope itself can be used for anchoring.
Pull the probe rope through a suitable anchoring
point, for example a welded eye, and fasten it
with two clamps.
The length of the loop will add to the transition
zone.The location of the clamps will determine
the beginning of the transition zone. The probe
length should be configured as the length from
the underside of the flange to the top clamp. See
section “Transition zones” on page 16 for further
information on Transition Zones.
Alternative chuck for flexible single lead
probes
Loosen the screws. Pull the probe rope through a
suitable anchoring point, for example a
welded-eye.
Tighten the screws. The required torque and hex
key dimensions:
4 mm wire: 15 Nm, 4 mm
6 mm wire: 25 Nm, 5 mm
Drain
Ø 0.3 in. (8 mm)